Product Profile of Liquid Flow Sensor with Analog Output
Liquid Flow Sensor with Analog Output is used to measure flow rates of virtually any fluid as low as 15ml/ minute or as high as 50 l / metre. Chandak Instruments Pvt. Ltd offers Liquid Flow Sensor with Analog Output. Chandak Instruments Pvt. Ltd also manufactures various types of flow meters. The Liquid Flow Sensor with Analog Output, Model U706, features a straight flow path. This Flow Sensor has PTFE, perfluoroelastomers,
And sapphire wetted parts which ensure compatibility with chemicals commonly found in microelectronics manufacturing processes, including de-ionized water, CMP slurries, acids, solvents etc. This Flow Sensor has
Pressure rating of 5.4 bars.
Product Specification of Liquid Flow Sensor with Analog Output
- Accuracy:±1.0% Full Scale (including linearity, best fit straight line)
- Calibration Interval:Calibration should typically be verified once every 12 months
- Compatible Liquids:Low viscosity (<10 cS recommended)
- Filtration:20 microns or less (Recommended)
- Output:4 - 20 mA Output Signal: Optional 4 mA at zero flow 500 ohm maximum loop resistance Not isolated
- Output VDC:0 - 10 VDC Output: Optional 0 VDC at zero flow 5 Kohm or greater output load Not isolated
0 - 5 VDC Output: Optional 0 VDC at zero flow 2.5 Kohm or greater output load Not isolated
- Power Requirements:12 - 15 VDC Units: 12 - 15 VDC, 50 mA typical
15 - 25 VDC Units: 15 - 25 VDC, 75 mA typical
22 - 25 VDC Units: 22 - 25 VDC, 50 mA typical
- Pressure Rating:80 psig (5.4 bar) working
100 psig (6.8 bar) overpressure
- Reliability:100,000 Hours MTBF
- Repeatability:±0.20% Full Scale
- Response Time:Typically <1 second for 97% of final value
- Temperature ratings:Environment: Operating: 0 to 50°C
Storage: 0 to 70°C
Fluid: Standard: 0 to 55°C
HT Suffix: 0 to 90°C