Product Profile of Rotameter Unit without Tubes
Anaesthetics India Pvt. Ltd. Is manufacturing Rotameter Unit without Tubes. The Rotameter Unit without Tubes has housing. The flow control is obtained by needle valves which can be easily distinguished from one another by touch & colour. Easy to read large size scales with fluorescent illumination, allow inside bobbins to be accurately read.
Anaesthetics India Pvt. Ltd. Is ISO 9001: 2000 & ISO 13485:2003 certified and also WHO GMP compliant company that offers Critical Care, Oxygen Therapy & Anaesthesia Equipments.
The operating principles of Rotameter Unit without Tubes are:
- Differential pressure āPā is created in the main Flow by providing an Orifice plate in the main pipeline.
- Because of this `P` a branch of flow moves through By-Pass line provided across the orifice plate from up stream side to down stream side of the Orifice Plate.
- Additional Range Orifice plate is provided in the By-Pass line which is designed such that flow through range Orifice plate flows in proportion with the flow through main orifice plate.
- Hence by measuring the flow from By Pass line the user can estimate the flow from Main Pipe line.
Models of Rotameter Unit without Tubes are:
- 3700 - Accomodates one 230mm tube.
- 3600 - Accomodates two 230mm tubes.
- 3500 - Accomodates three 230mm tubes